Przeglądanie stron 20
ULTIMA MOS 5 HART
17
Byte Format Description
12 Unsigned-8 Alarm Hi status : 0 – off, 1 – on , 2 – accepted
13 Unsigned-8 Alarm Lo status : 0 – off, 1 – on , 2 – accepted
14 Unsigned-8 Alarm Mid status : 0 – off, 1 – on , 2 – accepted
15 Unsigned-8 Power cycled flag
16 Unsigned-8 Event happened flag
17 Integer-8 % of FS
18-21 Integer-32 level – For all toxic gasses else = 0
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined
10.21 Command #164: Get Slow Changing Information
Request Data Bytes
Byte Format Description
None N/A N/A
Response Data Bytes
Byte Format Description
0 - 1 Signed-16 Reserved = 0
2 - 5 Float Supply Voltage
6 - 7 Signed-16 Sensor Voltage in miliVolt
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined