ULTIMA MOS 5 HART
18
10.22 Command #165: Get Set Up Information
This command sets the internal real-time clock.
Request Data Bytes
Byte Format Description
None N/A N/A
Response Data Bytes
Byte Format Description
0 Unsigned-8 Gas ID or sensor type
1 Enumerated Measured Units
2- 5 Unsigned-32 Full Scale
6 Unsigned-8 Alarm Hi level, % of FS
7 Unsigned-8 Alarm Hi Relay La/nL: 0 – nL, 1 - LA
8 Unsigned-8 Alarm Hi Relay En/dE: 0 – dE, 1 - En
9 Unsigned-8 Alarm Lo level, % of FS
10 Unsigned-8 Alarm Lo Relay La/nL: 0 – nL, 1 - LA
11 Unsigned-8 Alarm Lo Relay En/dE: 0 – dE, 1 - En
12 Unsigned-8 Alarm Mid level, % of FS
13 Unsigned-8 Alarm Mid Relay En/dE: 0 – dE, 1 - En
14 Unsigned-8
Alarm Mid Relay La/nL: 0 – nL, 1 - LA
15 Unsigned-8
Alarm delay
16 Unsigned-8
Sensitivity
17 Unsigned-8
Cal level, % of FS. Default is 50%
18 Unsigned-8
Cal IO type: 0 – Remote Cal line, 1 – manual solenoid, 2 –
automatic solenoid
19-20 Unsigned-16
Configuration Flags: Lock front panel flag, hazard watch, etc
21 Unsigned-8
Units on line
22 Unsigned-8
Number of votes
23 Unsigned-8
Sensor Life
24 Unsigned-8
Current Range: 0= 3.5 – 20, 1=1.25 - 20
Command-Specific Response Codes
Code Class Description
0 Success No Command-Specific Errors
1-127 Undefined